How to Clean Your Tuttnauer Sterilizer

How to Clean Your Tuttnauer Sterilizer

Maintaining your Tuttnauer sterilizer is essential for ensuring its effectiveness and extending its lifespan. Follow our guide below. This is for E and M machines ONLY

1. Prepare the Sterilizer

  • Open the unit and remove all trays
  • Ensure there's enough water in the reservoir by checking the dipstick or confirming the water level is half an inch below the safety valve

2. Add Chamber Brite

  • With the chamber empty, open a pack of Chamber Brite and pour it in a straight line down the center of the chamber, from the back to the front

3. Run a Cleaning Cycle

     For E Machines

  • Close the door and set and unwrapped cycle with NO drying time
  • ENSURE THE DRYING TIME IS SET TO 0
  • Press the start button to begin the cleaning process
  • Allow the cycle to complete

     for M Machines

  • Leave the door open and fill the chamber with water up to the groove as you would for a normal cycle 
  • Switch the multivalve to STE and close the door
  • Set the time for 25 minutes
  • Once the timer sounds, switch the multivalve to DRY/EXT and allow the chamber to depressurize
  • Switch the multivalve to 0

4. Drain the Reservoir

  • Once the cleaning cycle is complete, use the drain valve and a hose to empty the reservoir and remove the cleaning solution

5. Rinse and Reassemble

  • Refill the reservoir with distilled water and run another unwrapped cycle to rinse the chamber thoroughly
  • Drain the reservoir and wipe the chamber and water sensor
  • Reinsert the tray racks and trays

 

By Following These Steps, you will Maintain your Tuttnauer Sterilizer in Peak Condition