How to Clean Your Tuttnauer Sterilizer

Maintaining your Tuttnauer sterilizer is essential for ensuring its effectiveness and extending its lifespan. Follow our guide below. This is for E and M machines ONLY

1. Prepare the Sterilizer

  • Open the unit and remove all trays
  • Ensure there's enough water in the reservoir by checking the dipstick or confirming the water level is half an inch below the safety valve

2. Add Chamber Brite

  • With the chamber empty, open a pack of Chamber Brite and pour it in a straight line down the center of the chamber, from the back to the front

3. Run a Cleaning Cycle

     For E Machines

  • Close the door and set and unwrapped cycle with NO drying time
  • ENSURE THE DRYING TIME IS SET TO 0
  • Press the start button to begin the cleaning process
  • Allow the cycle to complete

     for M Machines

  • Leave the door open and fill the chamber with water up to the groove as you would for a normal cycle 
  • Switch the multivalve to STE and close the door
  • Set the time for 25 minutes
  • Once the timer sounds, switch the multivalve to DRY/EXT and allow the chamber to depressurize
  • Switch the multivalve to 0

4. Drain the Reservoir

  • Once the cleaning cycle is complete, use the drain valve and a hose to empty the reservoir and remove the cleaning solution

5. Rinse and Reassemble

  • Refill the reservoir with distilled water and run another unwrapped cycle to rinse the chamber thoroughly
  • Drain the reservoir and wipe the chamber and water sensor
  • Reinsert the tray racks and trays

 

By Following These Steps, you will Maintain your Tuttnauer Sterilizer in Peak Condition