ARE YOUR EXAM TABLES AND SCALES COMPLIANT?

New ADA Medical Diagnostic Equipment Accessibility Standards. DOJ Enforcement begins August 9th, 2026.

LEARN MORE

Download the ADA Compliance PDF

Project Checklist

  1. Seat Height

    Transfer surfaces shall be adjustable in height with:
    • A low transfer position at 17 inches maximum
    • A high transfer position at 25 inches minimum
    • At least 4 additional transfer positions in between
  2. Transfer Surface

    The seated transfer surface shall be:
    • 21 inches wide minimum
    • 17 inches deep minimum
  3. Transfer Supports

    • Gripping surface continuous and unobstructed
    • 15 inches long minimum
    • Diameter between 1.25 and 2 inches
    • Other measurement and height requirements
  4. Leg Supports

    Where stirrups are provided, leg support shall also be provided. Leg supports shall provide a method of supporting, positioning and securing the patient's legs
  5. Base Clearance

    The width of the base permitted within this clearance shall be 26 inches wide maximum at the edge of the examination surface.