ARE YOUR EXAM TABLES AND SCALES COMPLIANT?
New ADA Medical Diagnostic Equipment Accessibility Standards. DOJ Enforcement begins August 9th, 2026.
Download the ADA Compliance PDF
Project Checklist
-
Seat Height
Transfer surfaces shall be adjustable in height with:- A low transfer position at 17 inches maximum
- A high transfer position at 25 inches minimum
- At least 4 additional transfer positions in between
-
Transfer Surface
The seated transfer surface shall be:- 21 inches wide minimum
- 17 inches deep minimum
-
Transfer Supports
- Gripping surface continuous and unobstructed
- 15 inches long minimum
- Diameter between 1.25 and 2 inches
- Other measurement and height requirements
-
Leg Supports
Where stirrups are provided, leg support shall also be provided. Leg supports shall provide a method of supporting, positioning and securing the patient's legs -
Base Clearance
The width of the base permitted within this clearance shall be 26 inches wide maximum at the edge of the examination surface.